Answer:

It is acknowledged that the repeatability requirement of 0.2 % RSD is ambitious; it is nevertheless achievable under normal laboratory conditions. The error on the titre determination has to be limited as the content specifications of titrimetric assays are in most cases set at 100.0 ± 1.0 per cent.

...