Answer:

The limit of 2.5 mg specified in general chapter chapter 2.5.12 under “Suitability” under ‘Suitability’ concerns only the suitability of the procedure in this section for checking the solvent/reagent system. While performing this test, if your sample contains less than 2.5 mg of water, you can add a known amount quantity of water so that the total amount of water is within the indicated limits.